1. Summary
HHP developed a new prototype of an oscillating massage mattress. The accelerations
produced by this mattress is to be characterized in its working regimes, both in no-load and in
loaded circumstances, and to be checked with regulations.
2. Protocol
Eight accelerometers are used to measure the accelerations of the mattress: a tri-axial
accelerometer (PCB 356B21) and 3 uni-axial accelerometers (PCB 333B32) inside the
mattress (attached to the housing of a motor with unbalance), and 4 light-weight uni-axial
accelerometers (PCB 352A56) at the interface between the mattress and the test subjects.
The accelerometers are connected through an ICP signal conditioner (PCB 442C04) to a multifunction data acquisition card (NI 6251 PXI) in a PC running a custom build LabVIEW program. The data have been collected at an acquisition rate of 2 KHz during 40s. Care was
taken to separate the electrical circuits of the Massage Mattress and the Data Acquisitioning
Setup (ICP signal conditioner and PC).
For each vibrating zone of the mattress, 21 measurements were performed to describe the
vibration characteristics of the mattress for each combination of intensity (1 to 7) and added
weight (2, 4 or 6 kg). Additionally 7 measurements were performed with no weight added.
Furthermore 24 measurements were performed with 3 test subjects (60, 72 and 86 kg) at the 7 available intensities (gradually increasing the intensity of all zones, except the head zone, and
an additional measurement with the head zone at maximal intensity, noted in the raw result
tables as ‘intensity 8’).
To supply information about the rib and arm zone, the accelerometers of the head was moved
to the arm and the accelerometer of the feet was moved to the rib zone. An additional series
of 7 tests on a test subject of 72 kg was performed with this new configuration.
Data processing was done in Matlab by filtering the data (using either a low pass Butterworth
filter at 250Hz or the frequency weighting according to ISO2631-1), and subsequent
calculation of PSD (using a 32nd order Yule-Walker method for estimation of the frequency
components) and of the acceleration characteristics (root-mean-square, peak-to-peak, Crest
Factor, A(8), VDV, eVDV).
3. Results
The complete set of results is included in the appendix. In this section, some highlights of the
results are presented.
3.1 Zone characteristics
3.1.1 Frequency content

Figure 2: Base frequencies for the acceleration from the different zones (left: results with a load of 2kg; right:results with a subject of 72kg)
From the resemblance of the results with a load of 2 kg and with a subject of 72 kg (Figure 2),
it can be seen that the frequency content is independent of the load carried by the mattress.
Most of the vibrating zones produce acceleration signals with the same base frequency, except
the arms where a significant larger base frequency was found.
It is noted that the measured frequencies do not correspond with the frequencies that can be
read from the control box (20Hz at intensity 1 up to 80Hz at intensity 7). The frequency
content of the acceleration signals contains however multiple harmonic components. These
harmonic frequency components can sometimes even be stronger than the base frequency
itself (Figure 3).

Figure 3: Frequency content of the measured acceleration
along
the Z-axis at the waist (intensity 5)
3.1.2 Acceleration components
At the waist zone a tri-axial accelerometer is placed inside the mattress (attached to the
housing of a motor with unbalance). This allows us to look at the X, Y and Z component of
the acceleration produced at the waist zone.




Figure 4: Frequency components of the acceleration at the waist
(first row:
Z-axis inside the mattress;
second row: Y-axis inside the mattress; third row:
X-axis inside the mattress; last row: Z-axis on top
of the mattress)
From the results (Figure 4) it is clear that the motor mainly produces vibrations along the Zaxis
(vertical) and the Y-axis (longitudinal), in lesser amounts also along the X-axis (transversal). This is to be expected as there is no second unbalance to compensate the
vibrations in the Y-direction, and the unbalance is situated at one end of the housing (no
second unbalance at the other side of the housing), causing rotations of the housing, which in
turn result in vibrations along the X-axis). The results point out that there was sometimes a
problem with the wiring of the sensor underneath the test subjects, causing unreliable signals.
This was the case for subject 1 (60kg) at intensity 5, 6 and 7, and for subject 3 (86kg) at
intensity 1 to 4. The results are therefore not included in the presented graphs. Nevertheless it
can be seen that there is a lot of variation between the results of the test subjects. This
variation cannot be explained by the difference in total body mass alone, and must also be
dependent of body composition, differences in lying posture, …
4.1.3 Vertical accelerations
The vertical accelerations for the different zones are depicted in Figure 5 to Figure 16.

Figure 5: Vertical accelerations (RMS) for the head zone (weight tests)

Figure 6: Vertical accelerations (RMS) for the head zone (subject tests)
The signal from the head zone accelerometer revealed some abnormalities (high peaks in the signals, resulting in high crest factors) for Subject 3 (86 kg) in some cases (intensities 1, 4 and
7). These results were not included in the corresponding figure (Figure 6). For the head zone it seems that adding 2 to 6 kg is comparable with the head weight from the subject tests.

Figure 7: Vertical accelerations (RMS) for the shoulder zone (weight tests)

Figure 8: Vertical accelerations (RMS) for the shoulder zone (subject tests)
The results for the shoulder zone show large differences between the weight tests and the
subject tests. It seems as if the mattress experiences almost no effect of the weight of the
subject, as the RMS values of the subject tests are even higher than the results with 2 kg weight added. Furthermore it can be noted that the output vibration level drops drastically by adding 4 kg or more at the shoulder zone.
As already mentioned in the previous paragraph, the waist zone exhibits large differences in
vibration output for the different test subjects (although some of the experimental results had
to be excluded because of anomalies in the signal). More tests need to be carried out to clarify
the causes of this broad range in output.

Figure 9: Vertical accelerations (RMS) for the waist zone (weight tests)

Figure 10: Vertical accelerations (RMS) for the waist zone (subject tests)
The results for the feet zone reveal a peak in the acceleration output level (Figure 11 & Figure 12), which is probably due to resonance phenomena.

Figure 11: Vertical accelerations (RMS) for the feet zone (weight tests)

Figure 12: Vertical accelerations (RMS) for the feet zone (subject tests)
For the rib and the arm zone only one subject test (72 kg) is performed (as it involved a
change in accelerometer configuration).
For the rib zone, the results of the subject test are comparable to the addition of 2 kg (Figure
13 & Figure 14), while the arm zone is almost not influenced by the weight of the subject
(Figure 15 & Figure 16).

Figure 13: Vertical accelerations (RMS) for the rib zone (weight tests)

Figure 14: Vertical accelerations (RMS) for the rib zone (subject tests)

Figure 15: Vertical accelerations (RMS) for the arm zone (weight tests)

Figure 16: Vertical accelerations (RMS) for the arm zone (subject tests)
4.2 Safety regulations
4.2.1 Introduction
Vibration and shock can damage the human body. However, it is far from obvious what type
of damage will occur and what mechanisms are involved in the damage process, especially in
the case of persons in a standing, reclining or recumbent posture. It is therefore not possible to
state with any precision how the damage depends on the physical characteristics of the
vibration and shock, the characteristics of the exposed person, or other aspects of the
environment. Notwithstanding the incomplete understanding of the causes and nature of the
injuries produced by whole-body vibration (WBV), the belief that occupational and leisure
exposures to vibration can induce injury has led to various vibration standards.
The British Standard 6481 (1987) states that the primary quantity for expressing vibration
magnitude is the weighted root-mean-square acceleration (frequency weighting is done to
obtain a single value from multiple frequency or random vibration). However, it indicates that
r.m.s. magnitudes will underestimate some motions which are intermittent or contain
occasional high peak values, where the vibration dose value (VDV) must be used. As this is
not the case with the massage mattress, the r.m.s. method is used to evaluate the vibration
exposure.
ISO 2631 (1997) is applicable to motions transmitted to the human body as a whole through
supporting surfaces: the feet of a standing person, the buttocks, back and feet of a seated
person or the supporting area of a recumbent person. It applies primarily to seated persons,
since effects of vibration on the health of persons standing, reclining or recumbent are not
known. The guidance is applicable to vibration in the frequency range 0.5Hz to 80Hz which is
transmitted to the (seated) body as a whole (through the seat pan). It furthermore states that “The assessment of the vibration shall be made with respect to the highest frequencyweighted
acceleration determined in any axis on the seat pan. When vibration in two or more
axes is comparable, the vector sum is sometimes used to estimate health risk”.
Weighting and multiplying factors are not explicitly specified for recumbent persons in the
ISO 2631 (1997) standard for health, as the health effects of whole-body vibration in
recumbent posture are unknown. They are specified for comfort: Wk for vertical recumbent
(except head), Wd for horizontal recumbent and Wj for vertical recumbent (head). They are
also tentatively recommended in the BS 6481 as Wb (comparable to Wk from ISO2631) for
vertical recumbent and Wd for horizontal recumbent vibrations (it is assumed that the head is
never in direct contact with the full vibration magnitude).
The Member States of the European Union have agreed to the harmonization of certain
“social provisions”, including minimum standards for ensuring health and safety in
workplaces. In 2002 Directive 2002/44/EC was adopted. This is known as the Physical
Agents (Vibration) directive. It sets minimum requirements for the prevention of vibrationrelated
ill health. The daily exposure action and limit values in the Directive are all specified
as an 8-hour energy-equivalent frequency-weighted acceleration (known as A(8) value),
although vibration dose value alternatives are given for WBV.
The A(8) parameter is defined as:

The WBV daily exposure action value (EAV) is 0.5m/s² A(8) (or VDV 9.1m/s1.75); the WBV
daily exposure limit value (ELV) is 1.15 m/s² A(8) (or VDV 21m/s1.75). The daily exposure
times required to reach the EAV (or ELV) can be calculated from:
The times required to reach the ELV for the waist zone, the shoulder zone and the rib zone are
shown in Table 2, Table 3 and Table 4:
Table 2: Times [min] required to reach the ELV for the shoulder zone
Table 3: Times [min] required to reach the ELV for the rib zone

Table 4: Times [min] required to reach the ELV for the waist zone

It seems that the shoulder zone is the most limiting for the vibration exposure of test subjects,
with only 3 minutes before the action value is reached for subject 1 (60 kg). However, it must
be noted that, although they have the same body weight, there are huge differences between
the findings of subject 2 and 4. Therefore it is recommended that more subject tests are
carried out before general conclusions regarding the limiting aspect of the vibration exposure
can be drawn. Furthermore there is still a lot of uncertainty on the possible negative effects of
vibration on recumbent persons, resulting in lack of clarity in the legislation for the
application of the vibration exposure limits to recumbent subjects.
4. References
[1] British Standards Institution (1987) BS 6841. Measurement and evaluation of human
exposure to whole-body vibration.
[2] International Organization for Standardization (1997) ISO 2631-1. Mechanical
vibration and shock – evaluation of human exposure to whole-body vibration – part I:
general requirements.
[3] Griffin M.J., A comparison of standardized methods for predicting the hazards of
whole-body vibration and repeated shocks. Journal of Sound and Vibration (1998)
215(4), 883-914.
[4] European Parliament and the Council of the European Union (2002). Directive
2002/44/EC on the minimum health and safety requirements regarding the exposure of
workers to the risks arising from physical agents (vibration). Official Journal of the
European Communities, OJ L177, 6.7.2002, 13-9.
[5] Nelson CM and Brereton PF., The European Vibration Directive. Industrial Health
(2005) 43, 472-479.